Dendritics Gemscale-50 Owners Manual Page 3 | 
    
| 
 | 
    
![]()
  
    Place scale on 
  stable, flat, horizontal surface.
  
  
    Press Lever to open.
  
  
    Press Button to turn 
  on, Reads <188.88> 
  
  
    <CAL> Center 
  Calibration Weight on Pan.
  
  
    <50.00> Replace 
  weight and start weighing.
  
  
    Press Button once 
  briefly to zero Display.
  
  
    Press Button down 
  until Display shuts off or scale will automatically shut itself off in
  30 minutes.
  
  
    Battery requires 
  replacement when scale will not give a reading.
  
  
    Locate Battery 
  Compartment door on the bottom of the scale.
  
  
    Take a thin flat 
  object such as the flat end of stone tweezers. Put it into the battery door 
  groove and press inward and up. The battery door will flip open for you.
  
  
    Replace with a 9 Volt 
  Alkaline for 300 operating hours or a 9 Volt Heavy Duty for 200 operating 
  hours.
  
  
    Put the battery door 
  hinges in place. Press battery door down until it "clicks" and shuts securely.
  
  
    <Err> Weight too 
  heavy.
  
  
    <-Err> Weight below 
  zero, weighing pan missing.
  
  
    <----> Waiting for a 
  stable reading. Use draftshield.
  
  
    Unchanging error 
  indications for two minutes. Needs factory 
  re-calibration.
  
![]()
Images and webpage designs © 2001-2025 your webmaster, jb and Dendritics Inc. [-]